Development and validation of an alpha fetoprotein immunoassay
Abstract
Circulating alpha fetoprotein (AFP) is a diagnostic and prognostic biomarker for hepatocellular carcinoma (HCC) with potential utility as a pharmacodynamic endpoint in rodent tumor models. This application is limited, however, by low sample volumes, highlighting the need for sensitive, sample-sparing biomarker assay methods. In order to improve the utility of AFP as an oncology biomarker, we developed a method for AFP using the Gyrolab™, an automated microimmunoassay platform. Commercially available antibodies were screened to identify optimal combinations that were then used in a multi-factorial design of experiments (DOE) to optimize reaction conditions. Analytical validation included assessments of accuracy and precision (A&P), and dilutional linearity/hook effect, as well as reagent and sample stability. The method is reliable, with total error, a measure of accuracy and precision, less than 30% for all concentrations tested. AFP concentrations were measurable in diseased mice and undetectable in normal mice. Therefore, this novel, low volume AFP immunoassay is suitable for pre-clinical drug development, where its miniaturized format facilitates serial sampling in rodent models of cancer.